The S2000 Miniature
Fiber Optic Spectrometer from Ocean Optics was a low-cost, high-performance
spectrometer easily configured for thousands of UV-VIS-Shortwave NIR
applications from 200-1100 nm. The S2000 featured a high-sensitivity
linear CCD array that provided unusually high response and excellent
optical resolution in a miniature package.
Configuring the S2000 to work with external analog-to-digital
converters could sometimes be time consuming and complicated.
Several years ago we introduced USB-series spectrometers and the
S2000 ceased to be the most popular spectrometer choice.
See the
USB4000-series Plug-and-Play Spectrometer for details.
Miniature Optical Design
The S2000 features a familiar
optical design -- the crossed Czerny-Turner -- that has been reconfigured
for an optical bench small enough to fit into the palm of your hand. This
compact optical bench -- with no moving parts -- has a high-sensitivity
2048-element linear CCD-array detector that accepts light energy
transmitted through single-strand optical fiber and disperses it via a
fixed grating across the array. Despite its tiny size, the bench delivers
low stray light and high resolution, comparable to conventional 1/4-meter
monochromators.

The S2000 performance specifications
described here were achieved under the following conditions: The configuration
featured a 600 lines/mm grating (set for 300-950 nm), with a
tungsten-halogen light source for illumination and a 1000 µm optical fiber
for signal collection. The configuration did not include an entrance slit.
Integration times varied.
|
Dimensions:
|
141.6 mm
x 104.9 mm x 40.9 mm (with enclosure, master channel only)
|
|
Weight:
|
390 g (with
enclosure, master channel only)
|
|
Power
consumption:
|
110 mA @
5 VDC (master); 60 mA @ 5 VDC (slave)
|
|
Detector
range:
|
200-1100
nm
|
|
Detector:
|
2048-element
linear silicon CCD array
|
|
Gratings:
|
14 gratings: UV through Shortwave NIR
|
|
Entrance
aperture:
|
5, 10,
25, 50, 100 or 200 µm wide slits
|
|
Order
sorting filters:
|
Installed
bandpass and longpass filters
|
|
Focal
length:
|
42 mm
(input); 68 mm (output)
|
|
Optical
resolution:
|
~0.3-10.0 nm FWHM
(depending on grating and size of entrance aperture)
|
|
Stray
light:
|
<0.05% at 600 nm; <0.10% at 435
nm; <0.10% at 250 nm
|
|
Dynamic range:
|
2 x 10 8 (system); 2000:1
for a single scan
|
|
Sensitivity
(estimated):
|
86 photons/count; 2.9 x 10-17
joules/count; 2.9 x 10-17 watts/count (for 1-second
integration)
|
|
Fiber optic
connector:
|
SMA 905 to single-strand optical fiber
(0.22 NA)
|
|
Integration time:
|
3 milliseconds to 60 seconds with 1
MHz A/D card
2 milliseconds to 60 seconds with 2 MHz A/D card
|
|